- General Assay & Diagnostic Apparatus[10]
- Emergency & Clinics Apparatus[10]
- Clinical Analytical Instruments[7]
- Physical Therapy Equipments[7]
- Medical Consumables[10]
- Ultrasonic[10]
- Hospital Beds[3]
- Sterilization Equipments[2]
- Ward Nursing Equipments[3]
- Laser Beauty Equipment[7]
- Surgical Instruments[2]
- Education & Training[2]
- Contact Person : Mr. Hu Rooney
- Company Name : Shenzhen Hlongmed Biotech Co., Ltd.
- Tel : 86-755-86664839
- Fax : 86-755-86664933
- Address : Guangdong,Shenzhen,15-08, East Block,Yihai Plaza, Chuangye Road North, Nanshan
- Country/Region : China
disposable infusion pump
The Features of Dual Mode Infusion Pump HL-X801E
1. Two optional modes, drip-rate mode and volumetric mode.
2. Complies with IEC 60601, high safety designation with dual CPUs.
3. High brightness LCD display and user-friendly operation interface.
4. Easy switch among d/min, mL/h and volume/time units.
5. Available for Bolus mode.
6. Anti-reverse function on motor to prevent upstream.
7. Self-check when start up and during infusion.
8. Top pressure sensor for preventing underflow.
9. Pressure and air-in-line detection devices both with 8 sensitivities setting, dynamically display pressure value.
10. IV set management function: Up to 20 IV sets can be added, stored and deleted in memory, convenient for operation after calibration.
11. Adjustable 8 volume levels of alarm for convenient operation in different environments.
12. CE certified.
The Specifications of Infusion Pump HL-X801E
Infusion Mode | Drip-rate and volumetric |
Pumping Mechanism | Peristaltic finger pump |
Optional Infuse Rate Unit | d/min, mL/h and volume/time |
IV Sets | Standard IV sets of 10, 15, 20 and 60d/mL, with diameter 3.4 ~ 4.5mm. |
Flow Rate Range | 1.0 ~ 1200 mL/h |
Drip Rate Range | 10 d/mL IV set: 1.0 ~ 200 d/min 15 d/mL IV set: 1.0 ~ 300 d/min 20 d/mL IV set: 1.0 ~ 400 d/min 60 d/mL IV set: 1.0 ~ 150 d/min |
Increment | From 0.1 mL/h to 99.9 mL/h, 0.1 mL increment; ≥100 mL/h, 1 mL increment. |
Flow Rate Accuracy | ≤ ± 5% (after calibration) |
Drip Rate Accuracy | ≤ ± 1.5% |
Volume Limit | 1 ~ 9999 mL |
Volume Infused Display | 0 ~ 9999 mL & >9999 mL |
Purge Rate | 600 mL/h |
KVO Rate | 1 mL/h |
Bolus Mode | Available for Bolus mode, flow rate adjustable. |
Occlusion Pressure | Upstream and downstream pressure detection device with 8 sensitivities setting, 40 ~ 140KPa. |
Air-in-line Detection | Air bubble detected by ultrasonic and the device with 8 sensitivities setting, 50 ~ 1000μL. |
Alarm Volume | 8 levels setting |
IV Set Management and Flow Rate Calibration | IV sets can be added, stored and deleted in memory,convenient for operation after calibration. |
Anti-reverse Function | Anti-reverse function on motor to prevent upstream |
Self-check | Self-check when start up and during infusion |
Audio and Visual Alarms | Air-in-line, Door open, Upstream occlusion, Downstream occlusion, Infusion finished, No operation, Leakage, Bottom empty, Battery low, Pump error, Motor reverse, AC power lost, etc. |
Display | Blue LCD with high brightness |
Power Source | AC: 100V ~ 240V, 50Hz/60Hz, ≤ 25VA |
Battery | Rechargeable Ni-MH battery: DC12V, 2000mAh, life: ≥ 4 hours at 25mL/h, remaining battery displayed. |
Operating Conditions | Temperature: 5°C ~ 40°C Relative humidity: 20% ~ 90% Atmospheric pressure: 70 ~ 106KPa |
Storage Conditions | Temperature: -20°C ~ 55°C Relative humidity: 10% ~ 90% Atmospheric pressure: 70 ~ 106KPa |
Classification | Class 2, CF, IPX4 |
Dimensions and N.W. | 150mm x 150mm x 235mm; 2.3kg |
Application | Intravenous infusion for various liquid medicines, such as vascular expansion solution, anticancer drug, accelerative procreates drug, liquid nutrient, etc. Applicable for micro and rapid infusion. |
NOTE: The specifications subject to change without prior notice.
disposable infusion pump